30 100359-50 B, ©J AB 2006
LUCAS
I U G I
2.4 ADJUSTING LUCAS TO THE PATIENT
WARNING – INCORRECT HEIGHT
ADJUSTMENT
If the suction cup presses down too hard or too light
on the rib cage, the patient’s circulation will be
compromised.
Turn the ON/OFF knob to Adjust and adjust the
height of the suction cup immediately.
WARNING – INCORRECT POSITION
If the suction cup is not positioned correctly in
relation to the sternum, there is an increased risk of
damage to the rib cage and internal organs. In
addition, the patient’s circulation will be
compromised.
Make sure that the suction cup is positioned
according to instructions in Section 3.4 before
starting compressions.
If the position changes during operation,
immediately turn the ON/OFF knob to Adjust and
adjust the position of the suction cup.