Changing Setup Options
LIFEPAK 1000 Defibrillator Operating Instructions D-5
©2006-2012 Physio-Control, Inc.
Access Pulse Setup from the AED menu. The underlined bold options in Table D-7 are the
factory default settings.
Access the Manual menu from the Manual Mode option in Setup. The underlined bold options in
Table D- 8 are the factory default settings.
Table D-7 AED Setup Menu—Pulse Setup Submenu
Menu Item Help Message Options
PULSE CHECK
Enable pulse check prompt. NEVER: Never prompt for PULSE CHECK.
AFTER SECOND NSA: After every NSA
except for first analysis NSA result.
AFTER EVERY NSA: Only after NO SHOCK
ADVISED
.
ALWAYS: After every stack of shocks and
every NSA finding
.
PULSE PROMPT
Select prompt for patient vital
signs.
CHECK PULSE, CHECK BREATHING,
CHECK CIRCULATION, OPEN AIRWAY
.
AED
MONITORING
Enable monitoring while in
AED mode.
ON, OFF.
MONITORING
REPEAT
Select AED monitoring prompt
repeat time.
OFF, 1, 2, 3, or 5 minutes.
PREVIOUS PAGE
Go back to previous page.
Table D-8 Manual Setup Menu
Menu Item Help Message Options
MANUAL
ACCESS
Enable Manual mode access. ON, OFF.
ANALYZE
Enable SAS analysis in Manual
mode. (An
ANALYZE softkey is
provided in Manual mode.)
ON, OFF.
PREVIOUS PAGE
Go back to previous page.