Specifications
LIFEPAK 1000 Defibrillator Operating Instructions A-3
©2006-2012 Physio-Control, Inc.
Waveform
(continued)
With Infant/Child pads:
The following specifications apply from 25 to 175 ohms.
Energy Accuracy (into 50 ohms):
Selected energy 4 +/- 15%; 86 joules +/- 15% maximum
Rated energy output is the nominal delivered energy based on the energy
setting and patient impedance, as defined in the following chart.
* Energy setting selected
Rated Energy Output
20
30
40
50
60
70
80
90
25 50 75 100 125 150 175
Patient Impedance (ohms)
Energy (J
325 J *
300 J *
275 J *
250 J *
225 J *
200 J *
175 J *
150 J *
360 J *
Rated Energy Output
Patient Impedance ()
Energy (J)