LIFEPAK 1000 Defibrillator Operating Instructions A-5
©2006-2012 Physio-Control, Inc.
Electrical Protection: Input protected against high voltage defibrillator pulses per
IEC 60601-1. See Figure A-1.
Figure A-1 Defibrillation–protected, Type BF Patient Connection
Safety Classification: Internally powered equipment. IEC 60601-1
AED Mode
Shock Advisory
System (SAS):
ECG Analysis system that advises whether a shock is appropriate,
meets rhythm recognition criteria specified in DF80 and IEC
60601-2-4. In AED mode, the device allows a shock only if SAS
advises defibrillation.
Shock Ready Time Time to first shock (electrodes connected to a patient at power on
and initial rhythm finding is Shock Advised):
Less than 25 sec to 200 joules
Less than 30 sec to 360 joules
Energy Sequence: Multiple levels, configurable from 150 to 360 joules
Shock-to-Shock
cycle time (200J to
300J):
Less than 25 seconds
Time for a 3-shock
sequence
(200J/300J/360J):
Less than 70 seconds
Manual Mode
Energy Sequence Delivers energy at levels selected in Setup mode.
Charge Time Charge time:
200 joules in less than 7 sec (typical)
360 joules in less than 12 sec (typical)
ECG Mode
ECG Display Provides nondiagnostic ECG display of the patient’s heart rhythm.
Display
Size (Active viewing
area)
120 mm (4.7 in.) x 89 mm (3.5 in.)
Display Type 320 dot x 240 dot LCD with backlight
Frequency Response 0.55 Hz to 21 Hz (-3 dB), nominal
Waveform Sweep
Speed
25 mm/sec for ECG, nominal
Waveform viewing
time
Minimum 4 seconds