100057-00 F, ©JOLIFE AB 2007 25
LUCAS
TM
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE
6 Technical
specification
6.1 OPERATION
All pressure regulators supplied by your
distributor are approved for use with LUCAS.
Power source Air
Air
consumption
13.7 gallons/minute (52 litres/minute)
Air hose Permanently attached, 10.8 feet (3.3 m)
in length, with unique male connector.
Approved air
sources for use
with LUCAS
Pressure regulators or wall outlets for air
(medical grade/breathing quality) should
meet the following specifications:
1. Nominal supply pressure: 46-87 PSI
(3.17-6 bar).
2. Minimum required free flow rate in
relation to nominal pressure
according to Figure
2!. Example
shown: Nominal supply pressure 66.7
PSI (4.6 bar).
3. Pressure drop at minimum free flow
rate obtained from Figure
2!, must
not exceed the value obtained from
Figure
2@. The shaded area in
Figure
2# illustrates an unacceptably
high pressure drop for nominal supply
pressure 66.7 PSI (4.6 bar) and above.
The performance of the air supply
should be verified according to the
procedures in Section 6.1.1 and 6.1.2. If
you do not know how to do this, please
contact a clinical engineer or your
distributor, for help.
2!
2@
2#